Determination of selected trace elements in marine biota samples with the application of fast temperature programs and solid sampling continuous source high resolution atomic absorption spectroscopy: method validation |
| |
Authors: | Anna Maria Orani Petko Mandjukov |
| |
Affiliation: | 1. Department of Nuclear Sciences and Applications, Environment Laboratories, International Atomic Energy Agency, Principality of Monaco, Monaco;2. CNRS, IRD, Observatoire de la C?te d’Azur, Géoazur UMR 7329, Université de Nice Sophia Antipolis, Valbonne, France;3. Department of Chemistry, South West University “Neofit Rilski”, ul. “Ivan Mihaylov” 66, Blagoevgrad, Bulgaria |
| |
Abstract: | Analytical procedure for the determination of As, Cd, Cu, Ni, Co and Cr in marine biota samples using solid sampling high-resolution continuum source atomic absorption spectrometry (HR CS AAS) and accelerated fast temperature programmes has been developed. Calibration technique based on the use of solid certified reference materials similar to the nature of the analysed sample and statistics of regression analysis were applied. A validation approach in line with the requirements of ISO 17025 standard and Eurachem guidelines was followed. Accordingly, blanks, selectivity, calibration, linearity, working range, trueness, repeatability and reproducibility, limits of detection and quantification and expanded uncertainty for all investigated elements were assessed. The major contributors to the combined uncertainty of the analyte mass fractions were found to be the homogeneity of the samples and the microbalance precision. Traceability to the SI system of units of the obtained with the proposed analytical procedure results was also demonstrated. The potential of the proposed analytical procedure based on solid sampling HR CS AAS technique was demonstrated by direct analysis of marine reference biota samples. Overall, the use of solid sampling HR CS AAS permits obtaining significant advantages for the determination of selected trace elements in marine biota samples, such as straightforward calibration, a high sample throughput, sufficient precision, a suitable limit of detection and reduced risk of analyte loss and contamination. |
| |
Keywords: | Solid sampling high resolution continuum source atomic absorption spectroscopy fast programmes trace elements solid sampling marine biota calibration |
|
|