Continuous screening of analytical parameters facilitates efficient development of HPLC methods required for impurity profiling
Authors:Shuntaro Furukawa  Yutaka Hirakura  Kunikazu Moribe
Institution:1. Analytical Research Laboratories, Astellas Pharma Inc., Ibaraki, Japan;2. Graduate School of Pharmaceutical Sciences, Chiba University, Chiba, Japanshuntaro.furukawa@astellas.com;4. Graduate School of Pharmaceutical Sciences, Chiba University, Chiba, Japan
Abstract:Developing a robust analytical HPLC–UV method to characterize a drug candidate during an early stage of development is a major challenge when not all impurity standards are available. Here, we report our efforts to devise an efficient strategy for HPLC method development using continuous screening of analytical parameters without impurity standards. This strategy uses small incremental changes in the mobile phase pH and column temperature to trace each impurity on an overlay chromatogram. We tested this method using benzocaine as the active pharmaceutical ingredient (API), and compounds with similar structures to represent unknown impurities. Despite the coelution of peaks, results identified the number of impurities and indicated the starting point and parameter variables of the ensuing optimization step. Further, we demonstrated that the retention time of each peak as a function of mobile phase pH accounts for the apparent pKa of known and unknown compounds in the presence of an organic solvent. This information is critically important to the selection of a robust pH range for HPLC methods.
Keywords:Apparent pKa  high performance liquid chromatography  impurity profiling  method development  peak tracking  screening
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号