首页 | 本学科首页   官方微博 | 高级检索  
     


The Calibration Method of TOFr in the STAR Experiment
Authors:RUAN Li-Juan  WU Jian  DONG Xin  SHAO Ming  CHEN Hong-Fang  WANG Xiao-Lian  LI Cheng  HUANG Sheng-Li
Affiliation: Department of Modern Physics,University of Science and Technology of China,Hefei 230026,China
Abstract:a tray of time of flight detector based on mrPc technology took part in the physics run of star both in p+p and d+au collisions in 2003. together with the information from the time Projection chamber(tPc) and the pseudo–vertex Position detector (pvPd) of the star experiment,corrections are done to calibrate the new prototype tof. the intrinsic time resolution of tofr reaches 85ps and it extends the π/K and K/p separation up to 1.6gev/c and 3.0gev/c respectively.
Keywords:
点击此处可从《中国物理C(英文版)》浏览原始摘要信息
点击此处可从《中国物理C(英文版)》下载免费的PDF全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏