首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Self-referenced method for optical path difference calibration in low-coherence interferometry
Authors:Laubscher M  Froehly L  Karamata B  Salathé R P  Lasser T
Institution:Institute of Biomedical Imaging, Optics and Engineering, Swiss Federal Institute of Technology, CH-1015 Lausanne, Switzerland. markus.laubscher@epfl.ch
Abstract:A simple method for the calibration of optical path difference modulation in low-coherence interferometry is presented. Spectrally filtering a part of the detected interference signal results in a high-coherence signal that encodes the scan imperfections and permits their correction. The method is self-referenced in the sense that no secondary high-coherence light source is necessary. Using a spectrometer setup for spectral filtering allows for flexibility in both the choice of calibration wavelength and the maximum scan range. To demonstrate the method's usefulness, it is combined with a recently published digital spectral shaping technique to measure the thickness of a pellicle beam splitter with a white-light source.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏