Electrochemical detector with a 20 nl volume for micro-columns liquid chromatography |
| |
Authors: | K. Šlais D. Kouřilová |
| |
Affiliation: | (1) Institute of Analytical Chemistry, Czechoslovak Academy of Sciences, 611 42 Brno, Czechoslovakia |
| |
Abstract: | Summary A flow-through two electrode wall-jet cell with a platinum measuring electrode and a cell volume of 20 nl has been designed and evaluated. It has been used to detect phenols by reversed phase liquid chromatography using short micro bore columns. The linear dynamic range between the measured current and the concentration is greater than 103 (1.5×10–7–5×10–4 mol/l) and the minimum analyzable amount was found to be 10 pg for pyrocatechol. A negligible broadening in the detector permits the use of micro columns down to 0.5 mm internal diameter, packed with 5 m particles, without any substantial distortion of eluted zones.Presented at the 14th International Symposium on Chromatography London, September, 1982 |
| |
Keywords: | Liquid chromatography Short micro bore columns Small volume electrochemical detector |