首页 | 本学科首页   官方微博 | 高级检索  
     


Application of ZrC-coated tubes to the determination of aluminum in serum by graphite furnace atomic absorption spectroscopy
Authors:S. Costantini   R. Giordano  I. Vernillo
Affiliation:Laboratory of Applied Toxicology, Istituto Superiore di Sanità, Viale Regina Elena 299, 00100 Rome, Italy
Abstract:An improved method for the determination of aluminum in serum by atomic absorption spectrometry with a graphite furnace is presented. The method is suitable for the analysis of serum samples having normal and elevated aluminum content. The serum sample can be analyzed diluted or undiluted because the selected temperatures program minimizes the matrix interferences. High temperatures are made possible by the use of graphite Zr-coated tubes. The detection limit is 0.7 μg Al/liter. The coefficient of variation is 3.2% for serum concentration of 5 μg Al/liter and 1% for serum concentration of 50 μg Al/liter.
Keywords:
本文献已被 ScienceDirect 等数据库收录!