Magnetic Resonance Spectroscopy at 1.5 T with a Hybrid Metasurface
Authors:E A Brui  A V Shchelokova  A V Sokolov  A P Slobozhanyuk  A E Andreychenko  V A Fokin  P A Belov  I V Melchakova
Institution:1.University of Information Technology, Mechanics, and Optics, International Research Center Nanophotonics and Metamaterials,St. Petersburg,Russia;2.Federal Almazov North-West Medical Research Center,St. Petersburg,Russia;3.Department of Radiology,Kirov Military Medical Academy,St. Petersburg,Russia
Abstract:A hybrid metasurface realized by a double array of brass wires inserted into two high-permittivity dielectric slabs at both sides was used to perform a magnetic resonance spectroscopy experiment at a 1.5 T clinical magnetic resonance scanner. The metasurface coupled inductively to a transceive birdcage body coil located within the scanner’s bore. The metasurface demonstrated an enhancement of the signal-to-noise ratio of the magnetic resonance spectroscopy experiment in vitro. Up to a signal-to-noise ratio gain of 7.4 for choline and creatine spectral lines was observed in the presence of the metasurface compared to the body coil alone.
Keywords:
本文献已被 SpringerLink 等数据库收录!