首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Cobalt Hexacyanoferrate-Modified Electrode for Amperometric Assay of Hydrazine
Authors:Ravi Shankaran  D  Sriman Narayanan  S
Institution:(1) Department of Analytical Chemistry, University of Madras, Chennai –, 600 025
Abstract:A graphite electrode modified with cobalt hexacyanoferrate by mechanical immobilization was used for amperometric determination of hydrazine. The modified electrode exhibits good catalytic activity for the oxidation of hydrazine at a reduced overpotential with remarkable sensitivity. The modified electrode showed a linear response for hydrazine in the concentration range of 2.0 × 10–5 to 2.8 × 10–4 M. The detection limit was 9.8 × 10–6 M (S/N = 3). The proposed modified electrode was simple, sensitive, rapid, stable and promising.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏