首页 | 本学科首页   官方微博 | 高级检索  
     


Ultratrace determination of 1,2,4-trichlorobenzene in waste water by purge and trap/gas chromatography coupled to different detectors: Flame ionization (FID), electron capture (ECD), and multiple ion detection-mass spectrometry (MID-MS)
Authors:Raffaella Borrelli  Tiziana Fiorani  Piero Golfetto
Abstract:High resolution gas chromatography (HRGC) coupled to a “purge and trap” extraction-injection technique is described as a method of determining 1,2,4-trichlorobenzene in water at levels as low as parts-per-trillion (ppt). In order to investigate the interference from other volatile organic compounds (VOCs) several detection systems were compared: flame ionization, electron capture, and mass spectrometry-multiple ion detection. Concentrations ranging from 15 to 600 ng/L were analyzed in 20 ml standard aqueous solutions. The mean accuracy of the method varied from 89 to 103%, and its mean precision varied from ± 0.85 to ± 7.5 % for all detectors. The detection limits were 20 ng/L for FID, 2 ng/L for ECD and 0.5 ng/L for MID-MS detectors. The procedure was successfully applied to the analysis of industrial waste waters. The necessity of an appropriate internal standard to improve the quantitative determination and to determine possible losses or degradation of 1,2,4-trichlorobenzene during handling or storage is discussed.
Keywords:Capillary gas chromatography  Purge and trap extraction  1,2,4-TCB  Environmental analysis
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号