首页 | 本学科首页   官方微博 | 高级检索  
     检索      


On-line solid-phase extraction-thermal desorption-gas chromatography with ion trap detection tandem mass spectrometry for the analysis of microcontaminants in water
Authors:Arjan J H Louter  Jacco Van Doornmalen  Jolan J Vreuls  Udo A Th Brinkman
Abstract:An improved set-up for solid-phase extraction with thermal desorption coupled on-line to gas chromatography (SPETD-GC) is presented. It includes a newly designed liner for a programmable temperature vaporizer (PTV) and an improved water elimination system. The SPETD procedure now includes a washing step with HPLC-grade water to prevent degradation of analytes due to interaction with remaining sample constituents. The system was used to analyze surface and tap water samples over a 4-month period. No decrease of chromatographic or trace-enrichment performance was observed, and a liner packed with Tenax GR could be used for at least 150 analyses. The SPETD module was coupled to GC with ion-trap detection for mass spectrometric (MS) and MS/MS detection. The linearity and repeatability of the procedure for several pesticides which were tested in the 0.5–10 μg/1 range were fully satisfactory (1 μg/1, RSD range 5–11%; n = 5). When using sample volumes of 0.1 ml only, detection limits were as low as 0. 1-0.2 μg/1. As an example, the confirmation and quantification of a suspected pesticide in a real-life sample using electron impact and positive chemical ionization in both the MS and MS/MS mode is shown.
Keywords:Solid-phase extraction-thermal desorption  Gas chromatography  Ion trap detection  Tandem mass spectrometry  Water samples  Microcontaminants
设为首页 | 免责声明 | 关于勤云 | 加入收藏