首页 | 本学科首页   官方微博 | 高级检索  
     


High speed analysis of H2S and COS impurities in coal gas by capillary gas chromatography
Authors:J. B. Laurens  E. R. Rohwer
Abstract:There is a need for rapid, on-line, gas chromatographic analysis of trace levels of hydrogen sulfide and carbon oxysulfide in the oil-from-coal industry. In order to improve an existing method based on a polymer bed concentrator, a porous layer open tubular column, and a flame photometric detector, time-optimization of the various steps involved was attempted. Use of a redesigned concentrator, changed column operating conditions, and a photoionization detector reduced analysis times from 6.5 to 3 min and cycle times from 8 to 3 min. This could be achieved without compromising the 10 ppb detection limit required.
Keywords:High speed chromatography  Sulfur compounds  Photoionization detection  Trace analysis
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号