Colorimetric determination of iron in infant fortified formulas by sequential injection analysis |
| |
Authors: | A. N. Araújo Joan Gracia Jose´ L. F. C. Lima Manel Poch M. Lu´cia M. F. S. Saraiva |
| |
Affiliation: | (1) CEQUP/Dep. de Química–Física, Faculdade de Farmácia do Porto, P-4050 Porto, Portugal, PT;(2) Dep. Enginyeria Química, Universitat Autònoma de Barcelona, E-08193 Bellaterra, Spain, ES;(3) Laboratori d’Enginyeria Quimica i Ambiental, Universitat de Girona, E-17071 Girona, Spain, ES |
| |
Abstract: | A procedure is described for the colorimetric determination of iron in infant fortified formulas based on sequential injection analysis (SIA). Iron(III) complexation with thiocyanate is used as colour developing reaction. The system enables the determination of iron in the samples (after digestion by dry ashing and treatment with 0.2 mol/L nitric acid in the range of 0.50–20.0 mg/L, consuming 140 μL of the sample and 8 mg thiocyanate per determination. The reactor geometry and the adjustment of the ionic content of the calibration solutions is important for the accuracy of the results. A regression line according to the equation [Fe(III) (mg/L)]SIA=−0.3(±0.4)+1.03(±0.04) [Fe(III) (mg/L)]FAAS was obtained after comparative analysis of a set of 12 samples. The measurement rate was 34 s, thus allowing to analyze 100 samples per hour with a relative standard deviation lower than 2%. Received: 30 July 1996/Revised: 1 October 1996/Accepted: 4 October 1996 |
| |