首页 | 本学科首页   官方微博 | 高级检索  
     检索      


LC-MS/MS signal suppression effects in the analysis of pesticides in complex environmental matrices
Authors:B K Choi  D M Hercules  A I Gusev
Institution:Department of Chemistry, Vanderbilt University, Nashville, TN 37235, USA e-mail: Hercules@ctrvax.vanderbilt.edu, US
Rohm and Haas Company, Spring House, PA 19477, USA, US
Abstract:The application of LC separation and mobile phase additives in addressing LC-MS/MS matrix signal suppression effects for the analysis of pesticides in a complex environmental matrix was investigated. It was shown that signal suppression is most significant for analytes eluting early in the LC-MS analysis. Introduction of different buffers (e.g. ammonium formate, ammonium hydroxide, formic acid) into the LC mobile phase was effective in improving signal correlation between the matrix and standard samples. The signal improvement is dependent on buffer concentration as well as LC separation of the matrix components. The application of LC separation alone was not effective in addressing suppression effects when characterizing complex matrix samples. Overloading of the LC column by matrix components was found to significantly contribute to analyte-matrix co-elution and suppression of signal. This signal suppression effect can be efficiently compensated by 2D LC (LC-LC) separation techniques. The effectiveness of buffers and LC separation in improving signal correlation between standard and matrix samples is discussed.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号