Development of a Passive Sampler for Long-Term Measurements of Formaldehyde and Total Oxidants in Air |
| |
Authors: | G Bertoni C Ciuchini V Di Palo M Possanzini |
| |
Institution: | (1) Istituto sull Inquinamento Atmosferico del CNR A.d.R. di Roma, C.P. 10, Monterotondo Stazione, RM, Italy |
| |
Abstract: | A diffusive sampling device is described that is capable of reliable measurements of formaldehyde and total oxidants (Ox = ozone + nitrogen dioxide) at sub-ppbv concentration levels in ambient air. These species are collected on silica gel particles coated with 1-methyl-1-(2,4-dinitrophenyl)hydrazine (MDNPH) and phosphoric acid. The formaldehyde hydrazone (HCHO-MDNPh) and the N-methyl-2,4-dinitroaniline (MDNA) formed are extracted with acetonitrile and determined by HPLC with UV detection at 365ánm. The measured sampling rate for HCHO, 15.0 mLmin-1, agrees well with the theoretical value of 16.0, whilst an experimental sampling rate of 10.7 mLmin-1 (25% lower than the calculated one) is observed for Ox. The sampling rates seem to be independent of the sampling duration up to one month. The precision of the measurements for co-located passive samplers averaged is 7.3% for HCHO and 7.2% for Ox in urban air. |
| |
Keywords: | Column liquid chromatography Field validation Passive sampler Formaldehyde in air 1-methyl-1-(2 4-dinitrophenyl)hydrazine |