Determination of cadmium by flow injection-chemical vapor generation-atomic absorption spectrometry
Authors:Vargas-Razo C  Tyson J F
Institution:Department of Chemistry, University of Massachusetts, Amherst 01003-4510, USA.
Abstract:A method was developed for the generation of a "cold vapor" of cadmium by means of flow injection-chemical vapor generation from aqueous samples, the determination being conducted with an atomic absorption spectrometer (Pyrex glass T-cell). Several gas-liquid separator designs, atomizer designs, and the effect of several reagents previously reported as sensitivity enhancers (including cobalt, nickel, thiourea and didodecyl-dimethylammonium bromide) were investigated. The limit of detection, calculated as the concentration giving a signal equal to three times the standard deviation of the blank, was 16 ng L(-1), and the relative standard deviation was 1.4% for a concentration of 2 microg L(-1) and 3.8% for 0.1 microg L(-1). The addition of nickel and thiourea to the samples provided improved tolerance to the interference of coexisting ions. Two NIST certified reference materials, Montana Soil and Apple Leaves (respectively containing 41.7+/-0.25 mg kg(-1) Cd and 0.013+/-0.002 mg kg(-1) Cd) were accurately analyzed. The interference of lead was overcome by coprecipitation with barium sulfate, and the experimental values obtained were 41+/-1 mg kg(-1) Cd and 0.013+/-0.002 mg kg(-1) Cd, respectively.