首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Determination of Rutin in Drinks Using an Electrode Modified with Carbon Nanotubes-Prussian Blue
Authors:Edgar Nagles  Johisner Penagos-Llanos  Olimpo García-Beltrán  John Hurtado
Institution:1.Facultad de Ciencias Naturales y Matemáticas,Universidad de Ibagué,Ibagué,Colombia;2.Facultad de Ciencias Básicas,Universidad de la Amazonia,Florencia,Colombia;3.Departamento de Química,Universidad de los Andes,Bogotá,Colombia
Abstract:A new sensor was developed using a screen-printed carbon electrode modified with single-walled carbon nanotubes (SWCNTs) and Prussian blue (PB) coated with chitosan. The modified electrode allowed the oxidation and reduction of rutin at 0.25 and 0.096 V, respectively, with a ΔE of 0.154 V. Furthermore, the peak currents increase nearly 100% compared with the electrode without modification. The process was more reversible compared with the electrode modified with only SWCNTs or PB. Cyclic voltammetry was used to characterize the modified electrode surface. The quantification of rutin was more sensitive with adsorptive stripping voltammetry than with anodic stripping voltammetry. Adsorption potential, adsorption time and pH were optimized based on the oxidation of rutin: Eads =–0.10 V, tads = 60 s, pH 3.0. The detection limit (3σ/b) was 0.01 μM and the relative standard derivation was 3%. The new sensor was used in the quantification of rutin in black tea, coffee and synthetic drink of tea with satisfactory results.
Keywords:
本文献已被 SpringerLink 等数据库收录!