首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Voltammetric Determination of 8‐Nitroquinoline Using a Silver Solid Electrode and Its Application to Model Samples of Drinking and River Water
Authors:Tereza Rumlova  Jiri Barek  Frank‐Michael Matysik
Institution:1. Charles University in Prague, Faculty of Science, University Research Centre “Supramolecular Chemistry”, Department of Analytical Chemistry, UNESCO Laboratory of Environmental Electrochemistry, Albertov 6, CZ‐128 43 Prague 2, Czech Republic tel.:?+420?221951224;2. fax:?+420?224913538;3. University of Regensburg, Institute of Analytical Chemistry, Chemo‐ and Biosensors, Universit?tsstra?e 31, 93053 Regensburg, Germany
Abstract:This work is focused on the application of a silver solid electrode (AgE) for the development of modern voltammetric methods for the determination of submicromolar concentrations of biologically active compounds present in the environment. 8‐Nitroquinoline (8‐NQ), a well‐known chemical carcinogen, was chosen as a model substance. Differential pulse voltammetry (DPV) was used to study electrochemical behavior of 8‐NQ in different aqueous matrices. The following optimal conditions for determination of 8‐NQ in the concentration ranges from 2 to 100 µmol L?1 were used: Britton? Robinson (BR) buffer of pH 3.0, the regeneration potentials cycles (Ein=?1000 mV, Efin=?100 mV) and constant cleaning potential ?2000 mV. Practical applicability of AgE for the determination of micromolar concentrations of 8‐NQ was verified on model samples of drinking and river water.
Keywords:Silver solid electrode  Differential pulse voltammetry  8‐Nitroquinoline  Drinking water  River water
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号