Screen‐Printed Electrodes for Amperometric Determination of Iodide |
| |
Authors: | M. Khodari U. Bilitewski A. A. H. Basry |
| |
Affiliation: | 1. Department of Chemistry, Faculty of Science, South Valley University, Qena, Egypt;2. Department of Natural Products Biology, GBF‐Gesellschaft für Biotechnologische Forschung mbH, Mascheroder Weg 1, 38124 Braunschweig, Germany;3. Department of Pharmaceutical Chemistry, Faculty of Pharmacy and Pharmaceutical industries, Sinai University, El‐Arish, Egypt |
| |
Abstract: | ![]() An amperometric method for the determination of iodide ions has been developed using disposable, screen‐printed electrodes. The used sensors have a gold, graphite and platinum working electrodes with an area of about 7 mm2. Calibration curves exhibit a linear relationship between the electrode response and the iodide concentration up to 3.00 mM. The correlation coefficients for all calibration curves varied from 0.988 to 0.998. The relative standard deviations were equal to or less than 5.26 % (n=5). The lowest iodide concentration measured was 100 µM. |
| |
Keywords: | Screen‐printed electrodes Amperometry Iodide ions |
|
|