Improved Potentiometric and Optic Sensitivity of Polyaniline Film to Dissolved Oxygen by Incorporating Iron‐Porphyrin
Authors:Wei Liu  Yu Niu  Jorge P. Correia  Ana S. Viana  Gang Jin
Affiliation:1. NML, Institute of Mechanics, Chinese Academy of Sciences, 15 Bei‐si‐huan West Road, Beijing 100190, China;2. CQB, Faculdade de Ciências da Universidade de Lisboa, Edifício C8, Campo Grande, 1749‐016 Lisboa, Portugal
Abstract:Polyaniline (PAn) films can be used in an electrochemical and optical combination system to amplify the optical signal for dissolved oxygen (DO) detection. To further improve the sensitivity of PAn film to DO, an oxygen‐sensitive molecule, Fe(III) meso‐tetera (4‐sulfonatophenyl) porphyrin (FeTSPP) is incorporated into PAn during the film generation. Results show that, after incorporating FeTSPP into the PAn matrix, both optical and potentiometric responses to DO are improved. The optimal optical signal is obtained under the application of ?2.5 μA to the modified electrode. Under this applied current, both optical and potentiometric signals show linear relations with the DO concentration within the range of 0.00–4.63 mg L?1, and the sensitivities for optical and potentiometric signals are 4.18 grayscale units mg?1 L and 13.39 mV mg?1 L, respectively.
Keywords:Dissolved oxygen  Electrochemical and optical combination sensor  Fe(III) meso‐tetra (4‐sulfonatophenyl) porphyrin (FeTSPP)  Polyaniline  Total internal reflection imaging ellipsometry(TIRIE)