Matrix modification by Cu(NO3)2 for the determination of gold by electrothermal atomic absorption spectrometry with a molybdenum tube atomizer
Authors:Kiyohisa Ohta  Masayosi Yokoyama  Jinichi Ogawa  Takayuki Mizuno
Institution:(1) Department of Chemistry for Materials, Faculty of Engineering, Mie University, Tsu, 514 Mie, Japan
Abstract:Matrix modification by copper nitrate in electrothermal atomic absorption Spectrometry (ETAAS) of gold with a molybdenum tube atomizer has been investigated. The addition of copper nitrate served to eliminate the interferences from 104–105-fold concentrations of foreign elements at the 890 °C pyrolysis temperature. The absolute characteristic mass (giving 0.0044 absorbance) of gold in the presence of copper nitrate with the Mo atomizer was 0.26 pg and the detection limit was 38 pg/ml. These values were several times better than those obtained with graphite atomizers. The recovery of spiked gold in biological materials was in the range 96–106%. A sensitive and accurate ETAAS method was developed for complex matrix samples.
Keywords:electrothermal atomic absorption Spectrometry  gold  copper nitrate matrix modifier  molybdenum tube atomizer  biological materials
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏