Determination of naltrexone and 6 beta-naltrexol in plasma by high-performance liquid chromatography with coulometric detection
Authors:P Zuccaro  I Altieri  P Betto  R Pacifici  G Ricciarello  L A Pini  E Sternieri  S Pichini
Institution:Istituto Superiore di Sanità, Rome, Italy.
Abstract:A simple and reliable reversed-phase high-performance liquid chromatographic method with coulometric detection is described for the quantitation of naltrexone and its metabolite, 6 beta-naltrexol, in plasma samples of healthy volunteers who received orally 50 mg of naltrexone. The analytes and the internal standard, naloxone, are extracted with an octadecyl solid-phase extraction column before chromatography. The mobile phase is 0.01 M potassium phosphate (pH 3)-acetonitrile (85:15, v/v) and it is pumped at 0.8 ml/min. The coulometric detector is formed by two electrodes set at +0.20 V and +0.70 V, with a palladium reference electrode. The limit of quantitation observed was 5 ng/ml for both naltrexone and 6 beta-naltrexol. This method can be used to investigate pharmacokinetic parameters of different pharmaceutical preparations of this opioid antagonist.