首页 | 本学科首页   官方微博 | 高级检索  
     


A study of the effects of temperature and pressure on the retention time in series-coupled columns under conditions of constant mass flow rate
Authors:T. S. Buys  T. W. Smuts
Abstract:The technique of controlling chromatographic selectivity by the adjustment of column temperatures in systems of series-coupled columns is investigated by means of a general model incorporating the effects of temperature and mobile phase compressibility. In a previous article the performance of series-coupled columns was investigated under conditions of constant overall pressure drop. The present paper, on the other hand, considers systems in which a constant mass flow rate is maintained at all temperatures. The expressions derived for the effective partition coefficient and the retention time are compared with those of the previous paper. The numerical results for two-column systems exhibit the same major trends as those for constant pressure drop. It is shown that the effective partition coefficient is slightly dependent on the mass flow rate.
Keywords:Gas chromatography  Series-coupled columns with different stationary phases  Independent temperature control of columns  Constant mass flow rate  Retention time expressions  Theory  Selectivity control
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号