首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Capillary electrochromatography using fibers as stationary phases.
Authors:K Jinno  H Watanabe  Y Saito  T Takeichi
Institution:School of Materials Science, Toyohashi University of Technology, Japan. jinno@chrom.tutms.tut.ac.jp
Abstract:Fiber-packed capillary columns have been evaluated in chromatographic performance in capillary electrochromatography (CEC). The change of electroosmotic flow (EOF) velocity and selectivity using different kinds of fiber materials was examined. Although the EOF velocity among the different fiber packed columns was almost the same, retention of parabens was larger on the Kevlar-packed column than on the Zylon-packed one, and was larger on the as-span-type fiber-packed column than on the high-modulus-type packed one. Using 200 microm ID x 5 cm Kevlar packed column combined with a 100 microm ID x 20 cm precolumn capillary and a 530 microm ID x 45 cm postcolumn capillary, the separation of three parabens within 30 s was achieved. Other compounds were also separated in a few minutes by the fiber-packed CEC method.