A method of detecting capture γ-rays in a strong pulsed radiation field
引用本文:徐荣昆,唐正元,何锡钧.A method of detecting capture γ-rays in a strong pulsed radiation field[J].中国物理 B,2002,11(11):1145-1148.
作者姓名:徐荣昆  唐正元  何锡钧
作者单位:China Academy of Engineering Physics, Mianyang 621900, China;China Academy of Engineering Physics, Mianyang 621900, China;China Academy of Engineering Physics, Mianyang 621900, China
摘    要:We present a technology for diagnosing the D-T fusion process by detecting capture γ-rays. This technology provides an alternative route to diagnosing the D-T reaction process when a great deal of heavy Z materials surrounds the D-T region. A very important aspect of this paper is to focus on the methods of shielding low-energy γ-rays whose radiation intensity is 106 times higher than that of the capture γ-rays. Another aspect is about how to distinguish signal from noise. The result of a 50/1 signal-to-noise ratio indicates that the designed double-magnetic spectrograph is very successful for diagnosing the D-T fusion reaction process.

关 键 词:prompt  physical  process  diagnostic    D-T  fusion    capture  γ-rays    strong  pulsed  radiation  field
收稿时间:2001-12-03

A method of detecting capture γ-rays in a strong pulsed radiation field
Xu Rong-Kun,Tang Zheng-Yuan and He Xi-Jun.A method of detecting capture γ-rays in a strong pulsed radiation field[J].Chinese Physics B,2002,11(11):1145-1148.
Authors:Xu Rong-Kun  Tang Zheng-Yuan and He Xi-Jun
Institution:China Academy of Engineering Physics, Mianyang 621900, China
Abstract:We present a technology for diagnosing the D-T fusion process by detecting capture γ-rays. This technology provides an alternative route to diagnosing the D-T reaction process when a great deal of heavy Z materials surrounds the D-T region. A very important aspect of this paper is to focus on the methods of shielding low-energy γ-rays whose radiation intensity is 106 times higher than that of the capture γ-rays. Another aspect is about how to distinguish signal from noise. The result of a 50/1 signal-to-noise ratio indicates that the designed double-magnetic spectrograph is very successful for diagnosing the D-T fusion reaction process.
Keywords:prompt physical process diagnostic  D-T fusion  capture γ-rays  strong pulsed radiation field