A comparative study of234U/238U alpha-activity ratios determined by alpha-spectrometry and calculated from mass spectrometric data |
| |
Authors: | S. K. Aggarwal P. M. Shah |
| |
Affiliation: | (1) Fuel Chemistry Division, Bhabha Atomic Research Centre, Trombay, 400 085 Mumbai, India |
| |
Abstract: | 234U/238U α-activity ratios determined by α-spectrometry (AS) and those calculated from the atom ratio data using the half-life values are compared in some of the isotopic reference materials of uranium and a few other uranium samples. For α-spectrometry, electrodeposited sources were prepared and a large area passivated ion implanted (IPE) detector (450 mm2) was used for recording the α-spectra. The isotopic composition of U was determined by thermal ionisation mass spectrometry (TIMS) and the recommended half-life values of234U and238U were used to calculate the α-activity ratio. It is observed that234U/238U α-activity ratios calculated from the atom ratio data are consistently high, with a mean difference of about 5%, when compared to the α-spectrometry results. This discrepancy warrants confirmation by a few more laboratories and suggests redetermination of the half-life values of234U and238U. |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |
|