首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Evaluation of new high-frequency discharge lamps for atomic absorption and atomic fluorescence spectrometry of cadmium,lead and zinc
Authors:K Tsujii  K Kuga  S Murayama  M Yasuda
Institution:Central Research Laboratory, Hitachi Ltd., Kokubunji, TokyoJapan
Abstract:High-frequency discharge lamps with a hollow electrode are successfully utilized as the spectral line sources for atomic absorption and atomic fluorescence spectrometry of cadmium, lead and zinc. The sensitivities for atomic absorption spectrometry are superior to those obtained with commercially available hollow-cathode lamps by factors of 1.5 (Cd), 1.4 (Pb) and l.6 (Zn). Detection limits for non-dispersive atomic fluorescence spectrometry with graphite furnace atomization are 1 × 10-13 g (Cd), 3 × 10-11 g (Pb) and 2 × 10-13 g (Zn). The linear analytical range covers over four (Cd, Zn) and three (Pb) decades of concentration above the detection limits.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号