A Mechanism for Poor High Mass Performance in Fourier Transform Mass Spectrometry |
| |
Authors: | C. L. Holliman D. L. Rempel M. L. Gross |
| |
Affiliation: | 1. Midwest Center for Mass Spectrometry, Department of Chemistry, University of Nebraska, 68588 -0304, Lincoln, NE, USA
|
| |
Abstract: | The discovery of a mass-dependent electrical mechanism for signal loss of high mass ions in Fourier transform mass spectrometry is reported. Theoretical calculations and experimental evidence show the existence of resonances involving the z-motion and radial motion of the ion. The resonances are intrinsic in the cubic trap. This results in energy transfer between modes, expansion of the ion cloud, and a corresponding loss of signal. Quadrupolar trapping potential wells are proposed as a solution. |
| |
Keywords: | |
本文献已被 ScienceDirect SpringerLink 等数据库收录! |
|