Affiliation:(1) Analytical Department, DSM Research and Patents, PO Box 18, 6160 MD Geleen, The Netherlands;(2) Present address: Fundamental Polymer Research Department, DSM Research and Patents, PO Box 18, 6160 MD Geleen, The Netherlands;(3) Present address: Trainee of the Zuidlimburgse Laboratoriumschool Sittard, Montgomerystraat 7, 6135 BW Sittard, The Netherlands
Abstract:Summary Some quantitative implications of the use of microbore columns in column liquid chromatography are investigated. Although for several pumps the flow rate stability at 50 mgrl min–1 is slightly worse than that of pumps operating at 1 ml min–1, the quantitative performance with respect to repeatability and reproducibility of response factors equals that of conventional liquid chromatography. Thermostatting is strongly recommended for microbore column operation.
Keywords:Column liquid chromatography  Microbore column  Flow rate stability  Performance  Quantitative analysis