首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Monitoring of temperature variation in the focal region of an ultrasonic transducer
Authors:V G Andreev  A V Vedernikov  A V Morozov  V A Khokhlova
Institution:(1) Physical Faculty, Moscow State University, Vorob’evy Gory, Moscow, 119992, Russia
Abstract:A method for remote monitoring of temperature in the focal region of a high-intensity ultrasonic transducer is described. Results of measurements and theoretical simulation are presented. The measurements were conducted on a polymer sample with thermophysical and acoustic parameters close to the properties of a soft biological tissue. The sample was heated by a focused piezoelectric transducer with different values of radiation power. The delay of a probe pulse transmitted through the heated region perpendicularly to the axis of the intense ultrasonic beam was detected. The local character of temperature measurements was provided by focusing the probe pulse at the heated region. The application of an additional transducer installed confocally with the probing one provided an opportunity to enhance the precision of measurements. An analysis was conducted on the basis of a numerical solution of the heat conduction equation. The function of thermal sources in the heat conduction equation was calculated according to the results of measuring the pressure distribution in the focal region of the heating transducer. The experimental data obtained agree well with the results of simulation and demonstrate a fundamental possibility of using the proposed ultrasonic technique for remote temperature measurements.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏