首页 | 本学科首页   官方微博 | 高级检索  
     


A biocompatible needle-type glucose sensor based on platinum-electroplated carbon electrode
Authors:Chien-Yuan Chen  Eiichi Tamiya  Kazuhiko Ishihara  Yukio Kosugi  Yuan-Chi Su  Nobuo Nakabayashi  Isao Karube
Affiliation:Research Center for Advanced Science and Technology, University of Tokyo, Japan.
Abstract:A biocompatible needle-type glucose sensor with a 3-electrode configuration was constructed. A platinum-electroplated carbon stick was used as the working electrode, Ag/AgCl as the reference electrode, and a disposable hypodermic needle made of stainless steel as the counter electrode. A Nafion membrane, an immobilized glucose oxidase (GOD) membrane, and a biocompatible membrane with diffusion-limiting effect were coated successively onto the working electrode. The sensor showed a rapid response (< 120 s in batch operation), good reproducibility (RE < 3%), good stability (over 36 h in control serum), a wide dynamic range (5-600 mg/dL glucose), and superior biocompatibility. It was used to determine glucose in serum. The data obtained from the sensor showed good agreement with that from a clinical autoanalyzer (R > 0.95).
Keywords:  KeywordHeading"  >Index Entries Biocompatibility  2-methacryloyloxyethyl phosphorylcholine (MPC)   n-butylmethacrylate (BMA)  cellulose triacetate (CTA)  diffusion-limiting effect  glucose sensor  needle type  glucose oxidase  serum
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号