首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Enhanced mass resolution tandem mass spectrometry method for 2,3,7,8-tetrachlorodibenzo-p-dioxin detection with ion trap mass spectrometry using high damping gas pressure
Authors:Kemmochi Yukio  Tsutsumi Kaori  Nakazawa Hiroyuki
Institution:Water Environmental Technology Development Center, EBARA CORPORATION, 4-2-1 Honfujisawa, Fujisawa 251-8502, Japan. kemmochi10043@erc.ebara.co.jp
Abstract:Damping gas flow was optimized for 2,3,7,8-tetrachlorodibenzo-p-dioxin (TCDD) determination using ion trap mass spectrometer. A tandem mass spectrometry (MS-MS) method with better than unit-mass resolution (mass width, 0.3 u) was developed at a damping gas flow of 1.5 ml/min and a collision-induced dissociation (CID) voltage of 3.30 V. The relative standard deviation (R.S.D.) at the enhanced resolution was 2.9% in 24 h of consecutive injections. The detection limit was significantly improved because the efficiency of both precursor ion trapping and fragmentation increased with the damping gas flow. Product ion yield was 4.5 times higher and limit of detection was 3.2 times lower than at the default flow (0.3 ml/min and 1.65 V).
Keywords:Damping gas flow  Mass spectrometry  Polychlorinated dibenzodioxins
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号