On-column interactions in the direct GLC assay of hexetidine |
| |
Authors: | J Traveset V Such R Gonzalo E Gelpí |
| |
Institution: | (1) Departmento de Análisis y Control, Lacer S.A., Cerdeña 350, 25 Barcelona, Spain;(2) Instituto de Biofisica y Neurobiologia, CSIC, Avda. S. Antonio M. Claret, 171, 25 Barcelona, Spain |
| |
Abstract: | Summary Gas chromatography was used for the direct assay of hexetidine and triamine in a pharmaceutical formulation. During this work we came across a strong oncolumn interaction, in such a way that the GLC retention, selectivity and response parameters were seriously modified relative to the injections of hexetidine and triamine standard solutions. The interaction was traced to the coinjected hydrochloride form of a major component of the sample and in practical terms points to the potential quantitation errors that may be associated to similar systems. |
| |
Keywords: | Hexetidine Triamine Gas chromatography On-column interactions |
本文献已被 SpringerLink 等数据库收录! |
|