A novel microfluidic flow-injection analysis device with fluorescence detection for cation sensing. Application to potassium
Authors:Emilie Destandau  Jean-Pierre Lefèvre  Assia Chouai Fakhr Eddine  Serge Desportes  Marie Caroline Jullien  Rolland Hierle  Isabelle Leray  Bernard Valeur  Jacques Alexis Delaire
Institution:(1) D’Alembert Institute, ENS de Cachan, 61 avenue du Président Wilson, 94235 Cachan Cedex, France
Abstract:A microfabricated device has been developed for fluorimetric detection of potassium ions without previous separation. It is based on use of a fluorescent molecular sensor, calix–bodipy, specially designed to be sensitive to and selective for the target ion. The device is essentially made of a Y-shape microchannel moulded in PDMS fixed on a glass substrate. A passive mixer is used for mixing the reactant and the analyte. The optical detection arrangement uses two optical fibres, one for excitation by a light-emitting diode, the other for collection of the fluorescence. This system enabled the flow-injection analysis of the concentration of potassium ions in aqueous solutions with a detection limit of 0.5 mmol L−1 and without interference with sodium ions. A calibration plot was constructed using potassium standard solutions in the range 0–16 mmol L−1, and was used for the determination of the potassium content of a pharmaceutical pill. MediaObjects/216_2007_1132_Figa_HTML.jpg Figure Photography of the microfluidic channel showing the ridges in the PDMS substrate at the top of the channel
Keywords:μ  TAS  Fluorimetric detection  Potassium sensor  Flow-injection analysis  Passive mixer
本文献已被 PubMed SpringerLink 等数据库收录!