Authors:K Grob
Abstract:The sensitivity of an FID may change when the carrier gas flow rate changes during a chromatographic run. Sample parts which are eluted at reduced FID sensitivity produce a reduced peak area, hence are discriminated as compared to other components. Sensitivity changes were studied for hydrogen as carrier gas. For the detector tested, differences in the carrier gas flow rates of 1 ml/min shifted the FID sensitivity by 1 to 5% (depending on the fuel gas supply). Thus the stability of the sensitivity is no longer ensured as soon as the carrier gas flow rate is changed manually or by an automatic programmer during an analysis. Sensitivity drifts may also occur during temperature programmed runs with a pressure regulated carrier gas supply since the gas flow through the capillary drops with increasing temperature. Such shifts in the response became noticeable as soon as relatively high carrier gas flow rates combined with long range temperature programmes were used. The typical patterns of such discriminations are shown, closing with a discussion on the possibilities for minimizing such undesired effects.
Keywords:Gas chromatography  Capillary  Discrimination due to FID  Effect of carrier gas flow changes on detector sensitivity  Hydrogen as carrier gas