Abstract: | The study describes a simple and sensitive method for the sequential determination of hydrocortisone (HDC) and epinephrine (EP) in pharmaceutical formulations. The intrinsic fluorescence of epinephrine was measured before submitting the sample to a photochemical reaction to form a fluorescent product from HDC or HDC esters (succinate and acetate). The sample treatment eliminated the fluorescence from EP, thus minimizing spectral interference on the HDC fluorescence. Limits of detection (3Sb/m) were estimated to be 40ngmL–1 for HDC or HDC salts and 0.2ngmL–1 for EP. Recovery experiments, made with commercial and laboratory simulated formulations, gave values between 96 and 103% for both HDC and EP. This method can be easily automated, e.g. via flow-injection techniques. |