Modified INOvent for delivery of inhaled nitric oxide during cardiac MRI
Authors:Devendra Ganesh P  Hart Stephen A  Kim Yuli Y  Setser Randy M  Flamm Scott D  Krasuski Richard A
Affiliation:aDepartment of Cardiovascular Medicine, Heart and Vascular Institute, Cleveland, OH 44195, USA;bCleveland Clinic Lerner College of Medicine, Cleveland, OH 44195, USA;cChildren's Hospital Boston, Boston, MA 02115, USA;dDepartment of Radiology, Imaging Institute, Cleveland, OH 44195, USA
Abstract:

Background

The aim of this study was to assess the feasibility of delivering NO through a modified system to allow clearance of the magnetic field and thus compatibility with cardiac magnetic resonance (CMR). Nitric oxide (NO) is an inhalational, selective pulmonary vasodilator with a wide range of applications in a variety of disease states, including diseases that affect the right ventricle. Accurate assessment of dynamic changes in right ventricular function necessitates CMR; however, delivery of NO is only possible using equipment that is not magnetic resonance imaging (MRI) compatible (INOvent delivery system, Ohmeda, Inc., Madison, WI, USA).

Methods

The INOvent delivery system was modified by using 35 ft. of standard oxygen tubing to allow NO delivery through an electrical conduit and into the MRI suite. The concentrations of oxygen (O2), nitrogen dioxide (a harmful byproduct, NO2) and NO were measured in triplicate using the built-in electrochemical analyzer on the INOvent. After confirmation of safety, the system was used to administer drug to a patient x, and dynamic MRI measurements were performed.

Results

When the standard INOvent was set to administer 40 ppm of NO, the mean/standard deviation of gas delivered was as follows: NO: 42/0 ppm; NO2: 0.3/0.1 ppm; and O2: 93/0 ppm. In comparison, the gas delivery of the modified INOvent was follows: NO: 41/0 ppm; NO2: 0.5/0 ppm; and O2: 93.7/0.6 ppm. During administration to an index patient with severe pulmonic insufficiency (PI), a measurable reduction in PI was observed by CMR.

Conclusions

Nitric oxide can be administered through 35 ft. of standard oxygen tubing without significantly affecting dose delivery. This technique has potential application in patients with right-sided structural heart disease for determination of dynamic physiological changes.
Keywords:Modified INOvent   Inhaled nitric oxide   Cardiac MRI