Precise determination of bromine in PP resin pellet by instrumental neutron activation analysis using internal standardization |
| |
Authors: | Tsutomu Miura Ryo Okumura Yuto Iinuma Shun Sekimoto Koichi Takamiya Masaki Ohata Akiharu Hioki |
| |
Affiliation: | 1.National Metrology Institute of Japan,AIST,Tsukuba,Japan;2.Kyoto University Research Reactor Institute,Kyoto,Japan |
| |
Abstract: | Instrumental neutron activation analysis with the internal standardization was applied to the precise determination of Br in polypropylene resin of candidate certified reference material. The known amount of 197Au was used as an internal standard to compensate for neutron flux inhomogeneity, to improve the γ ray measurement uncertainty and the linearity of the calibration curves. The reliability of the proposed method validated using analytical results of BCR-681. The analytical result of Br in the sample was consistent with that obtained by ID-ICPMS. The relative expanded uncertainty (k = 2) was 1.5 %, and it was equivalent to that of ID-ICPMS. |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |
|