首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Certification by the Karl Fischer method of the water content in SRM 2890, Water Saturated 1-Octanol, and the analysis of associated interlaboratory bias in the measurement process
Authors:Margolis S A  Levenson M
Institution:National Institute of Standards and Technology, Analytical Chemistry Division, Gaithersburg, MD 20899-8392, USA. sam.margolis@nist.gov
Abstract:The calibration of Karl Fischer instruments and reagents and the compensation for instrumental bias are essential to the accurate measurement of trace levels of water in organic and inorganic chemicals. A stable, nonhygroscopic standard, Water Saturated Octanol, which is compatible with the Karl Fischer reagents, has been prepared. This material, Standard Reference Material (SRM) 2890, is homogeneous and is certified to contain 39.24 +/- 0.85 mg water/mL (expanded uncertainty) of solution (47.3 +/- 1.0 mg water/g solution, expanded uncertainty) at 21.5 degrees C. The solubility of water in -octanol has been shown to be nearly constant between 10 degrees C and 30 degrees C (i.e., within 1% of the value at 21.5 degrees C). The results of an interlaboratory comparison exercise illustrate the utility of SRM 2890 in assessing the accuracy and bias of Karl Fischer instruments and measurements.
Keywords:
本文献已被 PubMed SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号