首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Platinum particles-modified electrode for HPLC with pulsed amperometric detection of thiols in rat striatum
Authors:Cao Xu-Ni  Li Jin-Hua  Xu Hai-Hong  Lin Li  Xian Yue-Zhong  Yamamoto Katsunobu  Jin Li-Tong
Institution:Department of Chemistry, East China Normal University, Shanghai 200062, People's Republic of China.
Abstract:A new chemically modified electrode based on the immobilization of Pt particles is fabricated and exhibits electrocatalytic oxidation for L-cysteine (L-Cys), glutathione (GSH) and penicillamine (PEN) with relatively high sensitivity. It is also adaptable to HPLC for pulsed amperometric detection (PAD) of these thiols. PAD largely improves the detection sensitivity because the alternated polarizations can effectively clean and reactivate the electrode surface. It is shown that the peak currents of L-Cys, GSH and PEN are linear to their concentrations, with the calculated detection limit of 1.1 x 10(-7), 1.8 x 10(-7) and 3.8 x 10(-7) mol L(-1), respectively (S/N = 3). The method has been successfully applied to assess the contents of L-Cys and GSH in rat striatal microdialysates. The average contents of the two analytes in rat striatum are 2.6 x 10(-6) and 2.8 x 10(-6) mol L(-1), respectively.
Keywords:pulsed amperometric detection  Pt particles  l‐cysteine  glutathione  penicillamine  striatum
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏