ICP-MS法测定聚酯纤维中TiO2含量
引用本文:花金龙,姜逊,张玉莲,黄勇,岑良荧.ICP-MS法测定聚酯纤维中TiO2含量[J].化学分析计量,2011,20(2):30-32.
作者姓名:花金龙  姜逊  张玉莲  黄勇  岑良荧
作者单位:国家纺织品服装产品质量监督检验中心,广州,510220
基金项目:国家质检总局科技计划项目
摘    要:采用湿法和微波法对聚酯纤维样品进行消解,利用电感耦合等离子体质谱联用仪(ICP-MS)对样品中TiO2含量进行了测定,得到湿法消解样品测定值2 857.46μg/g(相对标准偏差为1.08%),微波消解样品测定值2599.21μg/g(相对标准偏差为2.76%).分别测定了两种消解方法的检出限,湿法消解检出限为4.0 ...

关 键 词:聚酯纤维  二氧化钛  湿法消解  微波消解

DETERMINATION OF TiO2 CONTENT IN POLYESTER FIBER BY ICP-MS
Hua Jinlong,Jiang Xun,Zhang Yulian,Huang Yong,Cen Liangying.DETERMINATION OF TiO2 CONTENT IN POLYESTER FIBER BY ICP-MS[J].Chemical Analysis And Meterage,2011,20(2):30-32.
Authors:Hua Jinlong  Jiang Xun  Zhang Yulian  Huang Yong  Cen Liangying
Institution:(National Textile and Garment Quality Supervision Testing Center, Guangzhou 510220, China)
Abstract:The polyester samples were digested by wet and microwave methods, and the results of sample reproducibility tested by ICP - MS were obtained. Measured values of wet digestion sample was 2857.46 μg/g ( RSD was 1.08% ) , and that of microwave digestion sample was 2 599.21 μg/g (RSD was 2.76% ). The detection limits of the two methods were obtained, the detection limit of wet digestion was 4.0 μg/g, and that of microwave digestion was 9.5 μg/g. The spiked recovery testing was carried out, the range of wet method was 84.82% -98.57% ( average value was 94.20% ) , and that of microwave method was 82.04% - 104.18% ( average value was 87.76% ). By comparison, the wet digestion method is more beneficial to digest the polyester samples.