首页 | 本学科首页   官方微博 | 高级检索  
     


On-line post-column fluorescence detection for N-terminal tyrosine-containing peptides in high-performance liquid chromatography
Authors:M Ohno  M Kai  Y Ohkura
Affiliation:Faculty of Pharmaceutical Sciences, Kyushu University, Fukuoka, Japan.
Abstract:A detection system based on on-line post-column fluorescence derivatization is described for the determination of N-terminal tyrosine-containing peptides by reversed-phase high-performance liquid chromatography. The peptides are automatically converted into fluorescent derivatives by reaction with hydroxylamine, cobalt (II) and borate after peptide separation on a reversed-phase column (TSKgel ODS-120T) followed by passage through an ultraviolet absorbance detector. The reaction system permits the fluorescence detection at 435 nm (emission) with excitation at 335 nm for N-terminal tyrosine-containing synthetic peptides in as little as picomole amounts. The facile fluorescence detection of N-terminal tyrosine-containing fragments produced from methionine-enkephalin by enzymatic degradation using a rat brain homogenate was achieved by comparison with the ultraviolet absorption detection at 215 nm.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏