首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Development and validation of a new high-performance liquid chromatography method for the determination of gliclazide and repaglinide in pharmaceutical formulations
Authors:Berecka Anna  Gumieniczek Anna  Hopkała Hanna
Institution:Medical University of Lublin, Faculty of Pharmacy, Department of Medicinal Chemistry, Jaczewskiego Str. 4, 20-090 Lublin, Poland.
Abstract:A new high-performance liquid chromatography method was developed and validated for the quantitation of gliclazide and repaglinide in pharmaceutical formulations. Determination was performed using a LiChroCART RP-18 column, a mobile phase containing acetonitrile-phosphate buffer (pH 2.1; 60 + 40, v/v), and ultraviolet (UV) detection at 225 nm. Repaglinide was used as an internal standard for gliclazide determination and gliclazide for repaglinide assay. The method was validated with respect to linearity, precision, robustness, ruggedness, accuracy, and specificity. The calibration graphs ranged from 0.015 to 0.09 mg/mL for gliclazide and 0.06 to 0.36 mg/mL for repaglinide. Intra- and interday relative standard deviation values for the standard solutions were 0.70 and 1.01% for gliclazide and 0.78 and 0.93% for repaglinide, respectively. Total recoveries of gliclazide and repaglinide from the laboratory-prepared mixtures were 99.82 +/- 0.58 and 101.50 +/- 0.46% for gliclazide and repaglinide, respectively mean +/- standard deviation (SD)]. In forced degradation studies, the effect of acid, base, oxidation, UV light, and temperature on both drugs was also investigated. Finally, the method was applied for the quality control of commercial gliclazide and repaglinide tablets. Total recovery was 100.40 +/- 0.35 and 104.46 +/- 0.23% for gliclazide and repaglinide, respectively (mean +/- SD).
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号