首页 | 本学科首页   官方微博 | 高级检索  
     


Detector Calibration and Measurement of Fundamental Parameters for X-Ray Spectrometry
Authors:Frank Scholze  Burkhard Beckhoff  Michael Kolbe  Michael Krumrey  Matthias Müller  Gerhard Ulm
Affiliation:Physikalisch-Technische Bundesanstalt, Abbestra?e 2-12, D-10587, Berlin, Germany
Abstract:Energy-dispersive X-ray spectrometry offers the opportunity for a fast and reliable determination of the composition of a specimen. For fundamental parameter based quantification, detailed knowledge of the X-ray generation cross sections and of the detection system’s efficiency is required. The detection efficiency is determined comparing the measured and calculated spectrum of undispersed synchrotron radiation (SR) from the electron storage ring BESSY II in the spectral range from 0.1 keV to 100 keV. Alternatively, monochromatized SR in the spectral range from 0.1 keV to 60 keV is used to determine the detection efficiency with a typical relative uncertainty of 1% to 2% by direct comparison with a reference detector. Employing well-calibrated detectors and monochromatized SR of well-known radiant power and high spectral purity, fluorescence yields have been investigated and resonant Raman scattering was studied as an example of a basic effect creating spectral background.
Keywords:: Quantitative X-ray spectrometry   energy-dispersive X-ray spectrometer   detector calibration   fundamental parameters.
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号