首页 | 本学科首页   官方微博 | 高级检索  
     


An anticoincidence-shielded Ge(Li) gamma-ray spectrometer and its application to radioanalytical chemistry problems
Authors:J. A. Cooper  L. A. Rancitelli  R. W. Perkins
Affiliation:(1) Battelle Memorial Institute, Pacific Northwest Laboratories, Richland, Washington, USA
Abstract:An anticoincidence-shielded Ge(Li) γ-ray spectrometer incorporating a largevolume, high-resolution coaxial diode inside a large plastic phosphor shield is described. It provides a peak-to-Compton ratio of 245: 1 for137Cs and a significantly improved sensitivity for both low- and high-level radionuclide measurements. Examples of its application to neutron activation analysis, analysis of proton activated sources, and environmental radionuclide analysis are illustrated. This paper is based on work performed under United States Atomic Energy Commission Contract AT(45-1)—1830.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号