Development and characterization of an all-solid-state potentiometric biosensor array microfluidic device for multiple ion analysis |
| |
Authors: | Liao Wei-Yin Weng Chen-Hsun Lee Gwo-Bin Chou Tse-Chuan |
| |
Institution: | Department of Chemical Engineering, National Cheng Kung University, Tainan 701, Taiwan. |
| |
Abstract: | A microfluidic device with an all-solid-state potentiometric biosensor array was developed using microfabrication technology. The sensor array included a pH indicator, and potassium and calcium ion-selective microelectrodes. The pH indicator was an iridium oxide thin film modified platinum microelectrode and the iridium oxide was deposited by an electrochemical method. The potassium and calcium ion-selective microelectrodes were platinum coated with silicon rubber based ion-selective membranes with respectively potassium (valinomycin) and calcium (ETH 1001) ionophores. The detection system was integrated with a micro-pneumatic pump which can continuously drive fluids into the microchannel through sensors at flow rates ranging from 52.4 microl min(-1) to 7.67 microl min(-1). The sensor array microfluidic device showed near-Nernstian responses with slopes of 62.62 mV +/- 2.5 mV pH(-1), 53.76 mV +/- 3 mV -logK+](-1) and 25.77 mV +/- 2 mV -logCa2+](-1) at 25 degrees C +/- 5 degrees C, and a linear response within the pH range of 2-10, with potassium and calcium concentrations between 0.1 M and 10(-6) M. In this study the device provided a convenient way to measure the concentration of hydrogen, potassium and calcium ions, which are important physiological parameters. |
| |