Verbesserte endpunktsbestimmung in der potentiometrischen analyse
Authors:FL Hahn
Institution:Durango 283, México 7, D.F. México
Abstract:If in a potentiometric titration each addition of reagent equals Δv and if the observed potential steps are, in decreasing order, ΔEm, ΔE1 and ΔE2, the ratio ρα = ΔE2E1is formed; ραΔv is the difference between the equivalence point and the known volume of reagent next to it. Both an extraordinary increase in the precision and a valuable simplification of the analysis are obtained. For Q = ΔEmΔE1 < 2.5 or ρα < 0.25 this value is approximated; but in this case a new series with doubled intervals may be formed from the potential steps observed in which ρα is very near to 0.5, found hence absolutely accurate.