Automatic coulometric titration of acids |
| |
Authors: | James J Lingane |
| |
Institution: | Department of Chemistry, Harvard University, Cambridge, Mass.U.S.A. |
| |
Abstract: | An apparatus is described for the automatic titration of acids by the constant current coulometric technique. The generator electrodes comprise a platinum cathode and a silver/silver bromide anode. The increase in pH resulting from the reduction of hydrogen ion at the cathode is indicated by a glass electrode, in conjunction with a Beckman Automatic Titrator which automatically monitors the titration and interrupts the generating current when the equivalence point pH is reached. Quantities of hydrochloric acid in the neighborhood of 0.12 millimole in 50 ml were titrated with a mean error of -0.07% and an average deviation from the mean of ±0.15%. The technique is applicable to any strong or weak acid, and to acid mixtures, provided that no substance is present which is either reducible at the platinum cathode or reactive in any way at the silver anode. |
| |
Keywords: | |
本文献已被 ScienceDirect 等数据库收录! |
|