首页 | 本学科首页   官方微博 | 高级检索  
     


A copper ion-selective electrode with high selectivity prepared by sol-gel and coated wire techniques
Authors:M.?Mazloum Ardakani  author-information"  >  author-information__contact u-icon-before"  >  mailto:mazloum@kashanu.ac.ir"   title="  mazloum@kashanu.ac.ir"   itemprop="  email"   data-track="  click"   data-track-action="  Email author"   data-track-label="  "  >Email author,M.?Salavati-Niasari,M.?Khayat Kashani,S.?M.?Ghoreishi
Affiliation:(1) Department of Chemistry, Faculty of Science, Kashan University, Kashan, Iran
Abstract:A sol-gel electrode and a coated wire ion-selective poly(vinyl chloride) membrane, based on thiosemicarbazone as a neutral carrier, were successfully developed for the detection of Cu (II) in aqueous solutions. The sol-gel electrode and coated electrode exhibited linear response with Nernstian slopes of 29.2 and 28.1 mV per decade respectively, within the copper ion concentration ranges 1.0×10–5–1.0×10–1 M and 6.0×10–6–1.0×10–1 M for coated and sol-gel sensors. The coated and sol-gel electrodes show detection limits of 3.0×10–6 and 6.0×10–6 M respectively. The electrodes exhibited good selectivities for a number of alkali, alkaline earth, transition and heavy metal ions. The proposed electrodes have response times ranging from 10–50 s to achieve a 95% steady potential for Cu2+ concentration. The electrodes are suitable for use in aqueous solutions over a wide pH range (4–7.5). Applications of these electrodes for the determination of copper in real samples, and as an indicator electrode for potentiometric titration of Cu2+ ion using EDTA, are reported. The lifetimes of the electrodes were tested over a period of six months to investigate their stability. No significant change in the performance of the sol-gel electrode was observed over this period, but after two months the coated wire copper-selective electrode exhibited a gradual decrease in the slope. The selectivity of the sol-gel electrode was found to be better than that of the coated wire copper-selective electrode. Based on these results, a novel sol-gel copper-selective electrode is proposed for the determination of copper, and applied to real sample assays.
Keywords:Sol-gel  Thiosemicarbazone  Ion-selective electrodes  Coated-wire electrodes  Copper
本文献已被 PubMed SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号