首页 | 本学科首页   官方微博 | 高级检索  
     检索      


High-performance liquid chromatographic analysis of basic drugs on silica columns using non-aqueous ionic eluents. I. Factors influencing retention, peak shape and detector response
Authors:R J Flanagan  I Jane
Abstract:The use of silica columns together with non-aqueous ionic eluents provides a stable yet flexible system for the high-performance liquid chromatographic analysis of basic drugs. At constant ionic strength, eluent pH influences retention via ionisation of surface silanols and protonation of basic analytes, pKa values indicating the pH of maximum retention. At constant pH, retention is proportional to the reciprocal of the eluent ionic strength for fully protonated analytes and quaternary ammonium compounds. The addition of water up to 10% (v/v) has little effect on retention if the protonation of the analytes is unaffected. Thus, it is likely that retention is mediated primarily via cation exchange with surface silanols. However, additional factors must play a part with compounds such as morphine which give tailing peaks at acidic or neutral eluent pHs.