首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Determination of arsenic, selenium, and antimony in cloud water by inductively coupled plasma mass spectrometry
Authors:Robert C Richter  Kamal Swami  Steven Chace and Liaquat Husain
Institution:(1) Wadsworth Center, New York State Department of Health, Albany, NY 12201-0509, USA, US;(2) Department of Environmental Health and Toxicology, School of Public Health, State University of New York, Albany, NY 12201-0509, USA, US
Abstract:The capability of inductively coupled plasma mass spectrometry in determining trace levels of As, Se, and Sb in cloud water was evaluated. Preliminary studies focused on identifying and eliminating potential interferences in the cloud water matrix, the choice of appropriate internal standards, and system optimization. The detection limits for As, Se, and Sb were 20, 100, 20 pg/mL using pneumatic nebulization, and 5, 25, 5 pg/mL, respectively, using ultrasonic nebulization with a precision of better than 5% RSD. The accuracy was demonstrated by the analysis of a NIST commercial reference material, SRM 1643d. In all cases, the results from ICP-MS analysis agreed within 4% of the certified values. Comparative analysis of cloud water samples obtained from a site downwind from large pollution sources (Whiteface Mountain, New York) and Changlagali Pakistan, a rural mountain peak, was carried out by hydride generation atomic absorption (HGAA) spectrometry. There was excellent agreement between the ICP-MS and HGAA results. Received: 31 July 1997 / Revised: 28 October 1997 / Accepted: 31 October 1997
Keywords:
本文献已被 SpringerLink 等数据库收录!